Software
Polwax Config – Datalogger Configuration Software
Polwax Config
Web-based configuration software for POLWAX single-use and multi-use dataloggers: recording, alarms, reports and start mode, in the browser.
- Platform
- Web application
- Loggers
- PTRANS, PTD50, PRH50, PTE50
- Languages
- EN, FR, DE, IT
Overview
Key features
- Runs in the browser with no installation required
- Temperature units, recording mode and interval
- Report format and start mode
- Alarm thresholds and time zone
- Password protection
- Interface in English, French, German and Italian
Technical specifications
| Supported loggers (web application) | PTE50, PTD50, PRH50, PTRANS-T10, PTRANS-TD10, PTRANS-T20, PTRANS-RH20 |
|---|---|
| Settings | Temperature units, recording mode and interval, report format, start mode, alarm thresholds, time zone, password protection |
| Languages | English, French, German, Italian |
Specifications are subject to change. Please check the current datasheet or ask our team before ordering.
Applications and industries
Application guides
-
Vaccine & medicine refrigerators
Log refrigerator temperatures at every interval with a reusable logger and produce PDF reports on connection.
-
Ultra-low temperature freezers
Record ultra-low freezer temperatures with only the probe inside the freezer and the logger outside.
-
Cold chain transport
Record shipment conditions in transit and generate a report automatically when connected by USB on arrival.
Application Demonstrations
Interactive examples of how this instrument is used. Choose an application, then try the Normal, Warning and Alarm conditions.
Pharma & life sciences
Vaccine & medicine refrigerators
Log refrigerator temperatures at every interval with a reusable logger and produce PDF reports on connection.
Illustrative simulation — not live instrument data.
Measurement points
- PTD50 logger on the middle shelf of the refrigerator
- PC: report opens on USB connection (PDF / CSV)
Monitoring workflow
- The PTD50 (temperature) or PRH50 (temperature and humidity) is configured in the Polwax Config web application.
- It records at an interval from 1 minute to 24 hours and stores up to 33,000 points.
- A green LED shows recording OK; a red LED shows an alarm. Up to 6 alarm limits can be programmed.
- Plugged into USB, the logger generates its PDF and/or CSV report automatically.
Products used
PTD50 / PRH50 Multi-Use Datalogger
PTD100 / PRH100 Multi-Use Mini Datalogger
PT70 / PRH70 Multi-Use Datalogger
Polwax Config – Datalogger Configuration Software This product
Alarm limits in this simulation are example settings. Set your own limits in Polwax Config.
Pharma & life sciences
Ultra-low temperature freezers
Record ultra-low freezer temperatures with only the probe inside the freezer and the logger outside.
Illustrative simulation — not live instrument data.
Measurement points
- Probe inside the freezer (cable through the door seal)
- PTE90 logger outside the freezer
Monitoring workflow
- The external probe is placed inside the freezer; the logger body stays outside.
- PTE50 measures down to −50 °C, PTE90 down to −90 °C (probe only), with ±0.4 °C accuracy.
- Readings are recorded at the set interval (default 5 minutes) with green / red LED status.
- The PDF / CSV report is generated when the logger is connected to USB.
Products used
PTE50 / PTE90 Ultra-Low Temperature Datalogger
Polwax Config – Datalogger Configuration Software This product
Freezer set-point and alarm limits shown are example settings.
Shipping & cold chain
Cold chain transport
Record shipment conditions in transit and generate a report automatically when connected by USB on arrival.
Illustrative simulation — not live instrument data.
Measurement points
- PTRANS logger packed with the goods
- Cross-dock stop (risk point)
- Clinic: plug the logger into USB
Monitoring workflow
- The single-use PTRANS logger is started with one button and packed with the goods.
- It records −30 to 70 °C (RH20 also 0–100 % RH) at the set interval; green LED for OK, red for alarm.
- For dry-ice shipments, PTRANS TD90 operates from −90 °C to 70 °C.
- On arrival, the logger is plugged into USB and the PDF / CSV report is available.
Products used
PTRANS Single-Use Datalogger
PTRANS TD90 Single-Use Dry Ice Datalogger
Polwax Config – Datalogger Configuration Software This product
Route, stops and alarm limits in this simulation are examples.
Documents and software
Related products

PTRANS-T10 / TD10 / T20 / RH20
PTRANS Single-Use Datalogger
Single-use USB temperature and temperature/humidity loggers for shipments. Press START to begin recording, then plug into…

PTD50 / PRH50
PTD50 / PRH50 Multi-Use Datalogger
Reusable USB dataloggers for temperature (PTD50) and temperature/humidity (PRH50) with display, up to 6 alarms, 33,000-point…

PTE50 / PTE90
PTE50 / PTE90 Ultra-Low Temperature Datalogger
Reusable USB dataloggers with external probe and display for ultra-low temperatures: PTE50 to -50 °C, PTE90…
- Image on request
CloudDASH360
CloudDASH360 Cloud Dashboard
Optional cloud dashboard for CLIMA WiFi, GSM and Ethernet dataloggers, showing data and alarms in a…
Enquiry
Ask about the Polwax Config – Datalogger Configuration Software
Tell us what you need to measure, the range and the installation. Our technical team will reply with a recommendation and quotation.